W-T2008-B haemodialysis machine is applicable for chronic renal failure and other blood Purification treatment.
This device should be used in medical units.
This device is specially designed, produced and sold for renal failure patients to receive haemodialysis, which is not allowed to be used for other purposes.
Haemodialysis, Isolated Ultrafiltration, Sequential Ultrafiltration, Hemoperfusion, etc.
Intelligent Double Operation System
LCD touch screen with button interface
Emergency power 30Mins (Optional)
Blood Pump
Spare Pump (for standby and also can be used for hemoperfustion)
Heparin Pump.
Hydraulic compartment(Balance chamber + UF pump)
Operation, Alarm information Memory function.
A/B Ceramic proportion pump , High precision, Corrosion-proof, Accuracy
Size & Weight Size: 380mm×400mm×1380mm (L*W*H)
Area: 500*520 mm
Weight: 88KG
Power Supply AC220V, 50Hz / 60Hz, 10A
Input power: 1500W
Back-up battery: 30 minutes (optional)
Water input pressure: 0.15 MPa ~0.6 MPa
21.75 P.S.I. ~87 P.S.I.
Water input temperature: 10℃~30
Working environment: temperature 10ºC ~30ºC at relative humidity of no more than 70%
Dialysate |
|
Dialysate temperature | preset range 34.0℃~39.0℃ |
Dialysate flux | 300~800 ml/min |
Dialysate concentration | 12.1 mS/cm ~16.0 ms/cm, ±0.1 ms/cm |
Dialysate mixing ratio | can set variety ratio. |
UF rate Flow range | 0 ml/h ~4000 ml/h |
Resolution ratio | 1ml |
Precision | ±30 ml/h |
Extracorporeal Part |
|
Venous pressure | -180 mmHg ~+600 mmHg, ±10 mmHg |
Arterial pressure | -380 mmHg ~+400 mmHg, ±10 mmHg |
TMP pressure | -180 mmHg ~+600 mmHg, ±20 mmHg |
Blood pump flow range | 20 ml/min ~400 ml/min(diameter:Ф6 mm) |
Spare pump flow range | 30 ml/min ~600 ml/min(diameter:Ф8 mm) |
Resolution ratio | 1 ml |
Precision | error range ±10ml or 10% of reading |
Heparin Pump |
|
Syringe size | 20, 30, 50 ml |
Flow range | 0 ml/h ~10 ml/h |
Resolution ratio | 0.1ml |
Precision | ±5% |
Sanitize |
|
1. Hot decalcification |
|
Time | about 20 minutes |
Temperature | 30~60℃, 500ml/min. |
2. Chemical disinfection |
|
Time | about 45 minutes |
Temperature | 30~40℃, 500ml/min. |
3. Heat disinfection |
|
Time | about 60 minutes |
Temperature | >85℃, 300ml/min. |
Storage Environment Storage temperature should be between 5℃~40℃, at relative humidity of no more than 80%. |
|
Monitoring System |
|
Dialysate temperature | preset range 34.0℃~39.0℃, ±0.5℃ |
Blood leak detection | Photochromic |
Alarm when erythrocyte specific volume is 0.32±0.02 or blood leak volume is equal or more than 1ml per liter of dialysate |
|
Bubble detection | ultrasonic |
Alarm when a single air bubble volume is more than 200µl at 200ml/min blood flow |
|
Conductivity | acoustic-optic, ±0.5% |
Optional Function | |
Blood pressure Monitor (BPM) | |
Display range Systole | 40-280 mmHg |
Diastole | 40-280 mmHg |
Accuracy | 1 mmHg |
Endotoxin filter — Dialysis fluid filter system | |
Balancing accuracy | ±0.1% of dialysate flow |
Bicarbonate holder | |
Concentrate | Bi-cart |
According to your actual needs, choose the most reasonable overall design and planning procedures