Hemodialysis Machine W-T2008-B HD Machine

Introduction

Device Name: Haemodialysis Machine (HD)

Class of MDR: IIb

Models: W-T2008-B

Configurations: the product is composed of circuit control system, monitoring system, blood extracorporeal circulation control system and hydraulic system, in which W-T6008S includes filter connector, replacement fluid connector, BPM and Bi-cart.

Intended use: W-T2008-B Hemodialysis Machine is used for HD dialysis treatment for adult patients with chronic renal failure in Medical Departments.

Product Details

Product Tags

Application Purpose of this Device

W-T2008-B haemodialysis machine is applicable for chronic renal failure and other blood Purification treatment.
This device should be used in medical units.
This device is specially designed, produced and sold for renal failure patients to receive haemodialysis, which is not allowed to be used for other purposes.

Forms of Therapy

Haemodialysis, Isolated Ultrafiltration, Sequential Ultrafiltration, Hemoperfusion, etc.

Features

 Intelligent Double Operation System
 LCD touch screen with button interface
 Emergency power 30Mins (Optional)
 Blood Pump
 Spare Pump (for standby and also can be used for hemoperfustion)
 Heparin Pump.
 Hydraulic compartment(Balance chamber + UF pump)
 Operation, Alarm information Memory function.
 A/B Ceramic proportion pump , High precision, Corrosion-proof, Accuracy

 Size & Weight Size: 380mm×400mm×1380mm (L*W*H)
 Area: 500*520 mm
 Weight: 88KG
 Power Supply AC220V, 50Hz / 60Hz, 10A
 Input power: 1500W
 Back-up battery: 30 minutes (optional)
 Water input pressure: 0.15 MPa ~0.6 MPa
 21.75 P.S.I. ~87 P.S.I.
 Water input temperature: 10℃~30
 Working environment: temperature 10ºC ~30ºC at relative humidity of no more than 70%

Parameter

Dialysate
Dialysate temperature preset range 34.0℃~39.0℃
Dialysate flux 300~800 ml/min
Dialysate concentration 12.1 mS/cm ~16.0 ms/cm, ±0.1 ms/cm
Dialysate mixing ratio can set variety ratio.
UF rate Flow range 0 ml/h ~4000 ml/h
Resolution ratio 1ml
Precision ±30 ml/h
Extracorporeal Part
Venous pressure -180 mmHg ~+600 mmHg, ±10 mmHg
Arterial pressure  -380 mmHg ~+400 mmHg, ±10 mmHg
TMP pressure  -180 mmHg ~+600 mmHg, ±20 mmHg
Blood pump flow range 20 ml/min ~400 ml/min(diameter:Ф6 mm)
Spare pump flow range 30 ml/min ~600 ml/min(diameter:Ф8 mm)
Resolution ratio 1 ml
Precision error range ±10ml or 10% of reading
Heparin Pump
Syringe size 20, 30, 50 ml
Flow range 0 ml/h ~10 ml/h
Resolution ratio 0.1ml
Precision ±5%
Sanitize
1. Hot decalcification
Time about 20 minutes
Temperature 30~60℃, 500ml/min.
2. Chemical disinfection
Time about 45 minutes
Temperature 30~40℃, 500ml/min.
3. Heat disinfection
Time about 60 minutes
Temperature >85℃, 300ml/min.
Storage Environment Storage temperature should be between 5℃~40℃, at relative humidity of no more than 80%.
Monitoring System
Dialysate temperature preset range 34.0℃~39.0℃, ±0.5℃
Blood leak detection Photochromic
Alarm when erythrocyte specific volume is 0.32±0.02 or blood leak volume is equal or more than 1ml per liter of dialysate
Bubble detection ultrasonic
Alarm when a single air bubble volume is more than 200µl at 200ml/min blood flow
Conductivity acoustic-optic, ±0.5%
Optional Function
Blood pressure Monitor (BPM)
Display range   Systole 40-280 mmHg
Diastole 40-280 mmHg
Accuracy 1 mmHg
Endotoxin filter — Dialysis fluid filter system
Balancing accuracy ±0.1% of dialysate flow
Bicarbonate holder
Concentrate Bi-cart

  • Previous:
  • Next:

  • Write your message here and send it to us

    Professional technical engineer dedicated to guide you

    According to your actual needs, choose the most reasonable overall design and planning procedures